The National Institute of Standards and Technology (NIST) provides a calibration service for power and energy meters to provide traceability for the over $400 billion annual sales of electricity in the US. Wideband power sensors are needed to modernize the NIST RF AC-DC measurement services and reduce calibration uncertainties at high frequencies associated with impedance mismatch effects.
The National Institute of Standards and Technology (NIST) has a requirement for 2 balance thermocouple sensor and is seeking information from sources that may be capable of providing a solution that will achieve the specifications described below:
The Contractor shall provide one (2) balance thermocouple sensor hat meet the following minimum specifications:
A. LIN 001: Balance thermocouple sensor for DC to 18 GHz
1. Shall contain two thermoelements, and can be used in a null arrangement
2. Shall have the frequency range DC to 18 GHz
3. Power range of -35 to +20 dBm
4. Connector N-type, male
5. Maximum input RF power: for CW signal: 316 mW/+25 dBm at 50 Ohm impedance and for Modulated Pulse Signal: 15 W/+42 dBm (2 microsecond)
6. Maximum input DC power: 316 mW/+25 dBm
7. Typical load resistance measured at the coaxial connector of 50 Ohms
8. Typical DC heater resistance measured at the DC connector of 50 Ohms
9. Thermocouple resistance: Single (RF side: 8 kOhms) and Dual (RF and DC: 16-17 kOhms
10. Thermocouple sensitivity for RF connector at 25 C: 1070 mV/W; DC Connector at 25 C: 1050 mV/W
11. Delivered with calibration with a calibration factor uncertainty of not more than 1.75 % at 18 Ghz
12. SWR of not more than 2.0 at highest frequency at the operating temperature
13. One unit with a 5 ft power sensor cable and the other unit with a 10 ft power sensor cable.
B. Balance thermocouple sensor for DC to 50 GHz
1. Shall contain two thermoelements, and can be used in a null arrangement
2. Shall have the frequency range DC to 50 GHz
3. Power range of -35 to +20 dBm
4. Connector 2.4 mm, male
5. Maximum input RF power: for CW signal= 316 mW/+25 dBm at 50 Ohm impedance and for Modulated Pulse Signal=15 W/+42 dBm (2 microsecond)
6. Maximum input DC power: 316 mW/+25 dBm
7. Typical load resistance measured at the coaxial connector of 50 Ohms
8. Typical DC heater resistance measured at the DC connector of about 97 Ohms
9. Thermocouple resistance: Single (RF side: 8 kOhms) and Dual (RF and DC: 16-17 kOhms
10. Thermocouple sensitivity for RF connector at 25 C: 1080 mV/W; DC Connector at 25 C: 2070 mV/W
11. Delivered with calibration with a calibration factor uncertainty of not more than 1.75 % at 50 Ghz
12. SWR of not more than 5.0 at highest frequency at the operating temperature
13. Must be operable with its own software (see Line 3 for specs)
14. Delivered with calibration with a calibration factor uncertainty of not more than 2.0 % at 18 Ghz
15. SWR of not more than 1.2 at highest frequency at the operating temperature
16. Must be operable with its own software (see Line 3 for specs)
C. Balance Thermocouple power sensor software
1. The software shall provide a interface for the devices specified in Line 001 and Line 002.
2. Shall work with Windows 11 and updateable with future Windows operating system versions.
3. The software can shall be used for the devices in Line 001 and Line 002 and transportable into different computers
HOW TO RESPOND TO THIS NOTICE
In responding to this notice, please DO NOT PROVIDE PROPRIETARY INFORMATION. Please include only the following information, readable in either Microsoft Word 365, Microsoft Excel 365, or .pdf format, in the response: Submit the response by email to the Primary Point of Contact and, if specified, to the Secondary Point of Contact listed in this notice as soon as possible, and preferably before the closing date and time of this notice. Please note that to be considered for award under any official solicitation, the entity must be registered and “active” in SAM at the time of solicitation response.
QUESTIONS REGARDING THIS NOTICE
Questions regarding this notice may be submitted via email to the Primary Point of Contact Junee Johnson junee.johnson@nist.gov listed in this notice. Questions should be submitted so that they are received 5 days prior to the response date. If the Contracting Officer determines that providing a written amendment to this notice to document question(s) received would benefit other potential respondents, the questions would be anonymized, and a written response to such question(s) would be provided via an amendment to this notice.
IMPORTANT NOTES
This notice is for market research purposes and should not be construed as a commitment by NIST to issue a solicitation or ultimately award a contract. There is no solicitation available at this time.
This notice does not obligate the Government to award a contract or otherwise pay for the information provided in response.
NIST reserves the right to use information provided by respondents for any purpose deemed necessary and legally appropriate.
Any organization responding to this notice should ensure that its response is complete and sufficiently detailed to allow the Government to determine the organization’s capability.
Respondents are advised that the Government is under no obligation to acknowledge receipt of the information received or provide feedback to respondents with respect to any information submitted.
After a review of the responses received, a synopsis and solicitation may be published on GSA’s eBuy or SAM.gov. However, responses to this notice will not be considered an adequate response to any such solicitation(s).
Thank you for taking the time to submit a response to this request.
Junee Johnson
DEPARTMENT OF COMMERCE
NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
DEPT OF COMMERCE NIST
DEPT OF COMMERCE NIST
ACQUISITION MANAGEMENT DIVISION
100 BUREAU DR.
GAITHERSBURG, MD, 20899
NAICS
Analytical Laboratory Instrument Manufacturing
PSC
LABORATORY EQUIPMENT AND SUPPLIES