Loading...
SSN_CVM-2026-134067
Response Deadline
May 15, 2026, 4:00 PM(EDT)4 days
Eligibility
Contract Type
Sources Sought
We expect to acquire seven new BSCs according to our specifications. The Contractor shall provide seven (7) biosafety cabinets consisting of:
• One (1) 6-foot wide unit
• Six (6) 4-foot wide units
All BSCs are expected to meet certification standards (NSF/ANSI 49) upon delivery.
Deliverables
The following are the requirements that should be met at the minimum:
• Units shall be available in unit widths of 6 ft (one unit) and 4 ft (six units)
• Units must meet the performance requirements of NSF/ANSI 49.
• Units shall have supply HEPA filters 99.99% efficient at a particle size of 0.3 micron by DOP test and shall be of full cabinet work zone width and depth.
• Units shall have exhaust HEPA filters 99.99% efficient at a particle size of 0.3 microns by DOP test
• Units shall be constructed from 16 or 18 gauge Type 304 stainless steel, forming a monolithic, sealed structure.
• Units shall possess a height adjustable base stand such that workspace height can be adjusted from at least 30 to 36 inches.
• Units’ power shall be supplied with a minimum 12-foot (2.5m) grounded electrical cable. Electrical supply should be 115 VAC, 60 Hz.
• Units shall be easily transportable through a 40 inch (1016 mm) wide, 83 in (2108 mm) tall door without disassembly.
• Units shall come standard with two outlets with drip proof covers; one gas valve/service coupling on right side wall; one service coupling on right side wall; two service couplings on left side wall.
• Units shall be capable of front filter removal without disassembly of the control panel and sliding window tracks/hardware.
• Units shall feature a sliding viewing window that operates smoothly from full closure to full opening at 20 inches.
• Units shall have a working access height of 10 inches.
• Units shall feature a non-metallic armrest/airfoil incorporating an at least 2 inch (51mm) forearm support area.
• Units shall feature an audible alarm and flashing LED to warn of an unsafe viewscreen position. The units will also signal power loss with visual and audible alarms
• Units shall include base stand with work surface height between 25 and 35 in
• Units shall include an ultraviolet (UV) light fixture.
• Units shall be warranted for parts and labor for at least 60 months from the date of acceptance.
• Pricing shall include shipping cost.
DEPARTMENT OF HEALTH AND HUMAN SERVICES
FOOD AND DRUG ADMINISTRATION
FDA OFFICE OF ACQ GRANT SVCS
FDA OFFICE OF ACQ GRANT SVCS
11601 Landsdown St
Floor 13
Rockville, MD, 20852
NAICS
Analytical Laboratory Instrument Manufacturing
PSC
LABORATORY EQUIPMENT AND SUPPLIES
Set-Aside
Total Small Business Set-Aside (FAR 19.5)