Sources Sought Synopsis Only
THIS REQUEST FOR INFORMATION (RFI)/SOURCES SOUGHT/RFI IS ISSUED SOLELY FOR INFORMATION AND PLANNING PURPOSES ONLY. THIS IS NOT A SOLICITATION.
THE SUBMISSION OF PRICING, CAPABILITIES FOR PLANNING PURPOSES, AND OTHER MARKET INFORMATION IS HIGHLY ENCOURAGED AND ALLOWED UNDER THIS RFI IN ACCORDANCE WITH (IAW) FAR 15.201(e).
The purpose of this synopsis is to gain knowledge of potential qualified sources and their size classification (HUBZone, 8(a), small, small disadvantage, veteran owned small business (VOSB), service-disabled veteran owned small business (SDVOSB), or large business relative to NAICS 333415 with size standards of 1250 employees. Note that if VA s pending market research establishes that two or more VOSB or SDVOSB, who are certified in U.S Small Business Administration site https://www.sba.gov can perform the project, VA will set aside the underlying solicitation per 38 U.S.C. §8127.Â
Responses to this synopsis will be used by the Government to make appropriate acquisition decisions. After review of the responses to this Sources Sought Synopsis, a solicitation announcement may be published on SAM, and/or GSA eBuy. Responses to this Sources Sought Synopsis are not considered adequate responses to the solicitation announcement.
General Requirements:
The Department of Veterans Affairs (VA), VA Long Beach Healthcare System, has a need for TSX Universal Series General Purpose Ultra Low Freezers. The Ultra - Low Freezers will be used to store biological samples for VA medical research studies.
The requirement is for brand name or equal: TSX Universal Series General Purpose Ultra-Low Freezer
Identify any domestic alternatives (meaning manufactured in the United States)
Nomenclature: Ultra-Low Temperature Freezer
Manufacture Information Fisher Healthcare/ Thermo Fisher Scientific
CLIN
Description
Qty
Unit of Measure
1
ULTRALOW GENERAL-PURPOSE FREEZER
2
EA
2
ENERGY SAVINGS REBATE
2
EA
3
POWER CORD
2
EA
4
SLIDINING DRAWER RACK
4
EA
REQUIREMENTS: The research lab requests the acquisition of ultra-low temperature freezers and additional equipment which will be suitable for the storage of biological samples. The Salient Characteristics must include:
- Ultra-low operating temperature
Operating temperature range must go down to 80C, preferably as low as 85C
Storage volume 19.31 CF
Temperature rating -40 to -86
Electrical power: 100V-230V 50/60 Hz
Instrument Rated Current: 8.9A 4.9A
Sound Pressure: 43.6 @ dBA -80C Setpoint
- Dimensions and Construction
Interior: Interior dimensions: 51.2 x 28.3 x 23.1
At least 3-4 fixed or adjustable shelves at least 12 h x 27 w with 1 increments
Shelf capacity: 150-165 lbs./shelf NTE 1,140 lbs. combined weight.
Storage capacity 400 standard 2 boxes
Separated interior doors for each shelf/compartment, to minimize cold air loss by opening only one interior compartment at a time
Insulation: Vacuum Insulation panels w/high-density water-blown polyurethane foam, electric heaters at breaker area.
Exterior: 78 x 37.4 x 28.3
Automatic alarms Adjustable warm/cold interior temperature alarms for defined maximum/minimum temperature, triggered when a defined delay time lapses, can also automatically forward the alarm to exterior monitor
Remote Outputs: RS485. 4-20 mA Output, Remote Alarm Dry Contacts
Mains/power outage alarm for interruption of power supply, along with battery back-up circuit
Battery alarm and other critical system alarms
Supports additional temperature probe / remote alarm interface for external temperature monitoring
Single outer door: Silicone-based gasket seal with 2 inner doors.
4x standard all-Direction casters with locks for stabilization and easy movement
Interface to support battery-operated back-up system
Locking mechanism: Handle key lock and Pad lock loop standard, HID controlled access
Backup System: LN or C02
- Electrical System
Soft-touch UI/ Main Circuit Breaker Poer Switch
Capacitive Touch Screen Input and Display with USB Data Retrieval at Eye Level controller type
High Temp Cutout Switch, Current, Logic protection Compressor Safeguard
Single RTD (1000-ohm Platinum RTD, class B, -196°C) controlled sensor
RS485, 4-20 mA Output, Remote Alarm Dry Contacts Remote Outputs
Adjustable in 1°C Increments adjustable warm/cold alarms
5V, POE Accessory Power
- Refrigeration Configuration
Two Stage Cascade System refrigeration system
Variable Speed Industrial Rated, Hermetically Sealed / 2 Compressor/number
938 W (max. speed) Compressor Capacity
Enhanced Micro-Channel and Forced-Air Cooled Condenser Type
Capillary Tube expansion device
Enhanced Cold Wall Design Evaporator type
Manual Defrost method
R290 / R170 + R290 Mix Refrigerant (1st/2nd stage)
Warranty: minimum 5-year standard warranty
Freight should be included (interior delivery)
Participating Healthcare Facilities:
VA Long Beach Healthcare System 5901 East 7th Street, Long Beach, CA; 90822-5201
In response to this announcement, please complete the information below:
Company:
Address:
Contact Name:
Phone No:
Email:
DUNS:
Business size information
Select all that applies:
Small Business
Emerging Small Business
Small Disadvantaged Business
Certified under Section 8(a) of the Small Business Act
HUBZone
Woman Owned
Certified Service-Disabled Veteran Owned Small in Vetbiz.gov
Certified Veteran Owned Small Business in Vetbiz.gov
Large Business
Other: ____________
All offered items must be listed on vendor FSS contract
FSS/ GSA Contract Holder of the requested item(s)
Yes No
FSS/ GSA Contract Number
Effective Date / Expiration Date
Does your company manufacture the requested product(s)? If no, answer the questions below.
Yes No
Does your company exceed the 500-employee alternative size standard for nonmanufacturers?
Yes No
Is your company primarily engaged in the retail or wholesale trade and normally sell the type of product(s) being supplied?
Yes No
Does your company take ownership or possession of the products(s) with its personnel, equipment, or facilities in a manner consistent with industry practice?
Yes No
Does your company supply the end product(s) of a small business manufacturer or processor made in the US, or obtain a waiver of such requirement?
Yes No
Identify any other companies owned (wholly or in-part) by the owners of this business who provide goods or services under the same or a related NAICS codes.
Not applicable
List:
Buy American Act Information
Are all items manufactured in the U.S? If not, list the items not manufactured in the US and the country of origin in the table provided below.
Yes No
Item #
Description
Country of Origin
Domestic Alternative Product
Do you have domestic alternatives? (Equal product that is manufactured within the United States) If so, please state the item, description and place of manufacture (country of origin) in the spaces below
Yes No
Item #
Description
Country of Origin
This notice is neither a request for competitive proposals or solicitation of offerors. This notice is to assist the VA in determining sources only and a solicitation is not currently available. If a solicitation is issued it will be announced at a later date, and all interested parties must respond to that solicitation announcement separately from this response. Responses to this Sources Sought is not a request to be added to a prospective bidders list or to receive a copy of the solicitation. Information received as a result of this notice will normally be considered solely for the purpose of determining whether to conduct competitive procurement. This notice does not represent a commitment by the Government to pay for costs incurred in the preparation and submission of information or any other costs incurred as a response to this announcement. Inquiries will only be accepted in writing via email to Rochelle.Malveaux@va.gov on or before Friday, May 22nd, 2026, by 2:00 pm PST Local Time.