Notice of intent to award a sole source contract to Fresenius USA Marketing, Inc. via SAC mandatory national contract no. 36C10G22D0012.
SOL: 36C24226Q0751
DUE: 07/03/2026
SET ASIDE: N/A
NAICS: 334516
POP STATE: 30 Days ARO
AGENCY: Department of Veterans Affairs; VA Healthcare System, 130 West Kingsbridge Road, Bronx, NY 10468.
Email: Jonatan.rondon@va.gov
The Department of Veterans Affairs, Margaret Cochran Corbin VA Campus, intends to award a one-time firm fixed-price, sole source contract to Fresenius USA Marketing, Inc. the authority of Federal Acquisition Regulation (FAR) 13.106-1(b) (1)(ii) to Fresenius USA Marketing, Inc. located at 920 Winter Street
Waltham, MA 02451 for their Portable Reverse Osmosis Water Purification System.
Background
This Statement of Work (SOW) establishes the requirements for the one-time procurement of four (4) brand name only AquaC 20 Amp portable reverse osmosis (RO) water purification systems with AAMI Water Profile Kit. The systems are required to support hemodialysis operations and to ensure a safe, AAMI-compliant water supply for patient care. The current portables require immediate replacement due to frequent, high-level bacterial contamination. Despite rigorous, in-depth disinfection and ongoing maintenance, the units advanced age and deterioration prevent effective sanitization.
Scope
Purchase of AquaC UNO RO 20 amp water purification system which includes equipment installation and training
Specific Tasks
The contractor shall furnish and deliver the following equipment:
Four (4) AquaC 20 Amp Portable Reverse Osmosis Systems (Product No. 24-0970-1)
Four (4) AAMI Water Profile Kits (Product No. 24-AAMI-0)
The contractor shall provide complete installation of all four (4) AquaC RO units, including mounting on existing Marcor carts, plumbing connections, and system commissioning.
The contractor shall provide on-site or training-center technical training for Medical Center clinical staff in accordance with the following parameters:
Service: AquaC System Tech Training at Training Center (Product No. 111122)
Initial training shall consist of a minimum of twenty-four (24) hours, not to exceed forty (40) hours, for up to five (5) healthcare professionals.
One (1) follow-up training session is included within the contract price.
Training must be initiated within thirty (30) calendar days after system installation; the follow-up training must be initiated within ninety (90) calendar days after installation. Failure to initiate within these windows shall result in expiration of included training, and additional fees will apply.
Additional attendees beyond five (5) and any additional training beyond initial plus one follow-up session are not included in this SOW and will incur additional charges.
AquaC Salient Characteristics:
System Type
Portable, single-patient reverse osmosis (RO) water purification system
Configuration
Modular cart-mounted unit; compact design for bedside/in-unit placement
Height
500 mm (1.64 ft) unit only, excluding cart
Width
520 mm (1.71 ft)
Depth
195 mm (0.64 ft)
Weight (Empty)
Approximately 35 kg (77.2 lb)
Weight (Full)
Approximately 37.5 kg (82.7 lb)
Cart Compatibility
Must be compatible with Marcor cart platform; designed for mounting on AquaC Cart US option
Portability
Must be cart-mounted and portable within the clinical environment; no permanent plumbing required
Water Quality Standard
Must meet or exceed ANSI/AAMI RD52 hemodialysis water quality standards for bacteria, endotoxins, and chemical contaminants
Permeate Output
Minimum 85 L/h (±10%) at 15°C (59°F) water inlet temperature
Bacterial / Endotoxin Rejection
Greater than 99% rejection rate for bacteria and endotoxins
Dissolved Salt Rejection
Greater than 96% average rejection rate for dissolved salts
Membrane Type
Spirally-wound RO membrane
Yield (Low Hardness)
Up to 70% at feed water hardness <1 °dH (<17.8 ppm CaCO3)
Yield (Higher Hardness)
Up to 60% at feed water hardness >1 °dH (>17.8 ppm CaCO3)
Concentrate Pressure (Max)
Maximum 290.1 psi
Noise Level (SUPPLY mode)
44 46 dB(A) measured at 1 meter distance
Permeate Output (Disinfection)
12 L/h (±10%) during ring heat disinfection cycle
Bacterial / Endotoxin Rejection
Greater than 99% rejection rate for bacteria and endotoxins
Dissolved Salt Rejection
Greater than 96% average rejection rate for dissolved salts
Membrane Type
Spirally-wound RO membrane
Yield (Low Hardness)
Up to 70% at feed water hardness <1 °dH (<17.8 ppm CaCO3)
Yield (Higher Hardness)
Up to 60% at feed water hardness >1 °dH (>17.8 ppm CaCO3)
Concentrate Pressure (Max)
Maximum 290.1 psi
Noise Level (SUPPLY mode)
44 46 dB(A) measured at 1 meter distance
Permeate Output (Disinfection)
12 L/h (±10%) during ring heat disinfection cycle
Bacterial / Endotoxin Rejection
Greater than 99% rejection rate for bacteria and endotoxins
Dissolved Salt Rejection
Greater than 96% average rejection rate for dissolved salts
Membrane Type
Spirally-wound RO membrane
Yield (Low Hardness)
Up to 70% at feed water hardness <1 °dH (<17.8 ppm CaCO3)
Yield (Higher Hardness)
Up to 60% at feed water hardness >1 °dH (>17.8 ppm CaCO3)
Concentrate Pressure (Max)
Maximum 290.1 psi
Noise Level (SUPPLY mode)
44 46 dB(A) measured at 1 meter distance
Permeate Output (Disinfection)
12 L/h (±10%) during ring heat disinfection cycle
Electrical Supply
120 V AC, 60 Hz, 20 Amp dedicated circuit
Nominal Power Variants
120 V 1000 VA or 120 V 1700 VA (per voltage variant)
Operating Mode
Continuous operation (Standby); SUPPLY mode used periodically during treatment
Electrical Safety Classification
Classified per EN 60601-1:2006 + A1:2013 / IEC 60601-1:2005 + A1:2012
Grounding Requirement
Must be connected to supply network with protective conductor (grounded outlet); no ungrounded connections permitted
Fuse Protection
Internal fuses for voltage component protection; replaceable only by trained/certified personnel
EMC Compliance
Compliant with applicable electromagnetic compatibility (EMC) standards; no stacked installation with other electrical devices
Interface / Connectivity
Fresenius hemodialysis device interface port; LAN port (RJ-45); compatible with Fresenius 2008T BlueStar dialysis machines
Thermal Disinfection Module
Module heat disinfection up to 82°C (179.6°F)
Thermal Disinfection Ring
Ring heat disinfection up to 90°C (194°F)
Membrane Service Life
Minimum expected life of 120 heat disinfection cycles
Chemical Disinfection
Compatible with peracetic acid solution (~0.2% effective concentration); system volume 2.5 L (0.66 gal)
Decalcification
Compatible with approximately 20% citric acid solution; automated decalcification program required
Rinse Interval
Automated rinse cycle: minimum every 4 hours; rinse volume 10 L (3 gal) or 10-minute rinse time
AutoSTART / AutoSTOP
System shall support automatic start in SUPPLY mode via communication cable connection to dialysis machine
Disinfection Validation
System must provide auditable disinfection reporting and daily operational data acquisition logs
Biocompatibility Standard
All components in contact with permeate must be biocompatible per DIN EN ISO 10993-1
Tubing / Piping Material
FEP, silicone, or stainless steel; food-grade with chemical resistance
Sensors
Titanium (Ti), Aluminum Oxide (Al O ), PBT
Module Tube
V4A stainless steel
Membrane Material
PSU/PA (polysulfone/polyamide)
Hydraulic Blocks / Valves
iPP, PVDF, Noryl GTX
Seals
EPDM (ethylene propylene diene monomer)
Sampling Ports
Integrated permeate sampling ports for microbiological and chemical analysis
Permeate Connection
Permeate nipple; optional permeate coupling; compatible with Fresenius hemodialysis machine water inlet
Feed / Concentrate Flow
Feed flow: 0.0 20 L/min (±10%); Concentrate flow: 0.0 20 L/min (±10%)
Permeate Consumption
0 6 L/min (±10%) during normal supply operation
Drain Connection
System drain must connect to available outlet; no recontamination of permeate circuit permitted
Water Supply Hose
Must connect to potable water supply meeting pre-treatment feed water requirements
Installation Type
Portable; no permanent wall or floor mounting required; Marcor cart-based
Stacking / Adjacency
System must NOT be stacked with other electrical devices; minimum separation distances from radiation sources required per IFU
Operational Qualification
Vendor must perform Installation and Operational Qualification (IQ/OQ) at time of installation
Display
Integrated digital display with operating mode status bar; multilingual support
Permeate Conductivity Monitoring
Continuous real-time permeate conductivity measurement and display; alarm on out-of-range condition
Feed Water Conductivity
Continuous real-time feed water conductivity measurement and display
Temperature Monitoring
Permeate and concentrate temperature display; heater temperature 1.0 135°C (±1°C); concentrate temperature 1.0 135°C (±1°C)
Rejection Rate Display
Real-time rejection rate display: 0.0 99% (±1%)
Effective Yield Display
Effective yield display: 0.0 70% (±5%)
Alarm System
Audible and visual alarms for: no water supply, fault drain, conductivity fault, sensor error, and other abnormal conditions; alarm must halt dialysis supply automatically
Operational Data Logging
System must support manual and automated daily operational data acquisition and reporting per AAMI requirements
Visual Status Indicator
Green visual indicator must illuminate to confirm system is ready for treatment
Place of Performance.
The location of delivery and installation will be:
James J. Peters VA Medical Center
130 West Kingsbridge Road
Bronx, NY 10468
Period of Performance.
Purchase of equipment includes one-year manufacturer warranty. Service and maintenance contracts are already in effect for the current equipment. Upon completion of installation of new equipment, the existing service contract will be amended to reflect new equipment.
This procurement is being conducted in accordance with FAR 13.106-1(b) (1)(ii), Soliciting from a Single Source. The NAICS code utilized for this procurement is 334516 , Analytical Laboratory Instrument Manufacturing with a Size Standard of 1,000 employees.
This Notice of Intent is not a request for Proposals or Quotations. No solicitation document exists, and no telephone or email inquiries will be accepted. No contract award will be made on the basis of proposals, or quotations received in response to this notice. Interested persons may identify their interest and capability to respond to the requirement and should furnish detailed data concerning their capabilities to provide services sought including pricing data and technical data sufficient enough to determine that a comparable source to the government is available. The government may consider all Expressions of Interest received prior to July 3rd, 2026. A determination by the government not to compete this proposed contract based upon responses to this notice is solely within the discretion of the government. Information received will be considered solely for the purpose of determining whether to conduct a competitive procurement. Responses to this notice must be emailed to jonatan.rondon@va.gov and received by 16:00 hours EST on July 3rd, 2026. Submitters are responsible for ensuring their EOI reaches the designated contracting officer on-time.