STATEMENT OF NEED
Zoll Zenix Series Advanced Defibrillators
INTRODUCTION/BACKGROUND:
Wilmington VAMC requires modern, reliable hospital defibrillator/monitors to support advanced resuscitation and patient transport within the facility. The ZOLL Zenix platform (or equal) provides multi-parameter monitoring, defibrillation, pacing, CPR feedback, and data integration capabilities required for patient care.
The following is being requested to be utilized by our two Rapid Response Teams in ICU and ED to respond to any code alerts within the main bldg. and outside bldg. areas. As of right now our outside bldgs. only have the capability to support AED only. With the Zoll Zenix series this is particularly suited for more advanced life support situation, and the compact size and weight make it easy for our Rapid Response Team to transport.
Currently we have Zoll brand defibrillators R series on our emergency crash carts and all the necessary supplies which staff are already trained on. The Zoll Zenix has the same defib connection pads as our R series.
PRODUCT/SPECIFICATIONS:
SCOPE:
The Department of Veterans Affairs, Wilmington VA Medical Center requests a brand name or equal procurement for the purchase of new Zoll Zenix Series Advanced Monitor/Defibrillator because it meets our needs for enhancing patient safety, accessibility, portability, and advanced monitoring capabilities.
This is a brand name or equal for the following: Zoll Zenix.
The following items are requested:
Line No.
Part No.
Description
Qty
001
8116-14-41-22331-00-
USA
Zenix Monitor/Defibrillator, Hospital, Ultimate
CCT Configuration + 12L 12-Lead, Pacing, SpO2, EtCO2, NIBP, IBP/Temp, Real CPR Help, Real BVM Help - Includes:
SurePower 4 Battery, Zenix Power Cord (NA), Zenix
MFC Cable, Zenix 12-lead ECG Cable (AAMI), Zenix
Accuvent Z-Link Cable, Masimo RD Rainbow SET
Patient Cable, Masimo RD SET Adult Reuseable
SpO2 Sensor, IRMA CO2 Analyzer, Zenix NIBP
Patient Hose, Zenix NIBP Cuff (Reusable, Adult
Plus), Zenix IBP Z-Link Cable (Abbott), Zenix
Temperature Z-Link Cable, Zenix, Z-Fold Paper,
Zenix MFC Self Test Plug Kit and Five (5)-year
Hospital warranty.
Parameter Details: Real CPR Help - Dashboard
display of CPR Depth and Rate for Adult and
Pediatric patients, Visual and audio prompts to coach
CPR depth (Adult patient only), Release bar to
ensure adequate release off the chest, Metronome to
coach rate for Adult and Pediatric patients. See-Thru
® CPR artifact filtering 12-Lead ECG with Inovise
Interpretative Algorithm (Full 12 ECG lead view with
both dynamic and static 12-lead mode display. 12-
Lead ECG cable - includes 4-Lead limb lead cable
and removable precordial 6-Lead set) ZOLL
Noninvasive Pacing Technology Real BVM Help:
Dashboard provides real-time ventilation feedback on
both volume and rate for intubated and non-intubated
patients. AccuVent Z-Link Cable included. (Accuvent
disposable sensors sold separately) SunTech NIBP
technology. 10 foot single lumen hose and SunTech
Reusable Adult Plus Cuff included Two
Temperature monitoring channels with digital
displays. One Temperature Z-Link cable included.
Temperature probes sold separately
2
002
8016-000111-01
SurePower 4 Battery
1
003
8016-001000
SurePower 4 Charger Adapter
004
8000-001128
AccuVent Sensors, Box of 10
5
005
8000-010109
Masimo IRMA Airway Adapter Adult, Box of 25
5
006
8016-000120-01
Zenix OneStep Connector
2
007
8016-000860-08
Zenix Carry Case, Slim
2
008
8012-0206
12-lead ECG Simulator
1
009
8000-000878-01
Zenix Fan Fold Paper, Grid, Box of 10
5
Required specifications/all equivalents must meet the following:
General
Weight: 13.2 lbs. (5.99 kg) with battery
Dimensions: 9.4 in (23.9 cm) x 10.3 in (26.2 cm) x 8.6 in (21.8 cm)
Enclosure Protection: Solid foreign object: EN 60529, IP3X Water: EN 60529, IP3X
Operating Temperature: 32 to 104°F (0 to 40°C)
Transport and Storage Temperature: -4 to 41°F (-20 to 5°C)
Humidity: 15 to 95% (non-condensing)
Vibration: 60068-2-64
Shock: 60068-2-27, 50g, 6ms half sine
Drop: IEC 60068-2-31, 18 in (0.5 meters) functional drop 60601, 78 in (2 meter) safety drop
Altitude: -1,276 to 15,000 feet
(-389 to 4,572 meters)
Monitor/Display
Size: 8.4 in (21 cm) diagonal
Type: Color LCD 800 x 600 touchscreen
Display: Up to 4 waveforms
Defibrillator
Waveform: Rectilinear Biphasic waveform
Patient Population: Adult, pediatric, neonatal/infant
Patient Impedance Range:15 to 300 ohms (hands-free electrodes or external paddles) 7 to 300 ohms (internal paddles)
Energy Selections: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 85, 100, 120, 150, 200 joules
Charge Time: Less than 7 seconds with a new, fully charged battery at 77°F (25°C) for 15 successive charges
Rapid Shock: Shock/No Shock decision with pre-charge of the device in 4 to 5 seconds
Advisory Defibrillation: Shock Advisory Function prompts rescuers through ECG analysis, shocks and CPR throughout a rescue event. The ECG is analyzed with shock/no shock decision provided to rescuer.
Pacer
Type: External transcutaneous pacing
Pulse Types: Rectilinear, constant current
Pulse Width: 40 ms +/- 2 ms
Pacer Rate: 30 - 180 PPM (+/- 1.5% for Fixed Mode)
Output Current: 10 - 140 mA +/- 5% or 5 mA (whichever is greater)
Minimum Pacer Output: 10 mA
Modes: Demand, Fixed
Output Protection: Fully defibrillator protected and isolated
Cardioversion
Synchronized defibrillator pulse to patients' R wave.
"SYNC" message displayed on monitor and markers shown on both monitor and recorded ECG.
Real-Time Clinical Feedback
See-Thru CPR: See-Thru CPR: Removes compression-related artifact from the ECG via an adaptive filtering technique.
Real CPR Help: Real-time depth and rate feedback.
CPR Dashboard: Numeric readout of pediatric and adult depth and rate, release indicator, perfusion performance indicator (PPI).
Real BVM Help: Provides ventilation feedback including rate and tidal volume in ml
ECG Monitoring
Input: 3-lead, 5-lead, 12-lead cables, paddles, therapy electrodes with MFC cable
Lead Selections: Paddles, Pads, I, II, III, AVR, AVL, AVF, V1-V6.
Heart Rate Range:20 - 300 beats/minute
Heart Rate Accuracy:3 BPM or +/- 3%, whichever is greater
ECG Size Range:0.25, 0.5, 1, 2, 3, 4 cm/mV
ECG Sweep Speed: User selectable25 mm/sec or 50 mm/sec
External Patient Monitor Connections
ECG Output: 1.0V/cm of deflection on recorder; <25ms delay from patient
CO2
Patient Population: Adult, pediatric, neonatal/infant
ET Tube Diameter: Single Patient Use (SPU),SPU Adult/pediatric: >4.0mm
SPU Pediatric/neonatal/infant: 4.0mm
Accuracy for all Conditions: +/-(0.3 kPa + 4% readings
Respiration Rate: 0 to 150 +/- 1 breaths/minute
Warm-Up Time: < 10 seconds (concentrations reported and full accuracy)
Non-Invasive Blood Pressure
Patient Population: Adult, pediatric
Method: Oscillometry
Control: Automatic and manual measurement
Auto Mode Intervals: 1, 2, 3, 5, 10, 15, 30, 60 mins
Masimo RD SET SpO
SpO2 Saturation Range: 0 - 100%
SpO2 Saturation Accuracy:
No Motion:
Adult/pediatric/infant: 70 - 100%, +/-2%
Neonate: 70 - 100%, +/- 3%
Motion: 70 - 100%, +/- 3% adults/pediatrics/infants/neonates
SpO2 Pulse Rate Range: 25 - 240 beats/minute
Masimo RD Rainbow Set SET
Masimo rainbow SET SpCO:
Range: 0 99%
Accuracy: 1 40%, ± 3%,
Adults/pediatrics/infants
Masimo rainbow SET SpMet:
Range: 0 99%
Accuracy: 1 15%, ± 1%,
Adults/pediatrics/infants/neonates
Masimo rainbow SET SpHb:
Range: 0 25 g/dL, 0-15.5 mmol/L,0-250 g/L
Accuracy: 8-17 g/dL, 1 g/dL adults/pediatric
Masimo rainbow SET SpOC: Range: 0 to 35 ml/dL
Masimo Perfusion Index (PI): Range: 0.02% - 20%
Masimo Pleth Variability Index (PVI): Range: 0.02% - 100%
Patient Log Files
Full Disclosure Data Storage:
Contains continuous waveform data, treatment markers, CPR data, 12-lead snapshots, monitor snapshots, automatic snapshots, trend records and vital signs numeric updates.
Full Disclosure Storage Capacity:
Up to 150 patient cases with FIFO storage or to the maximum storage capability
Printer / Recorder
Type: High resolution thermal array
Annotation: Snapshots, alarms, and treatment summaries
Paper Width: 80 mm
Paper Speed: 25mm and 50mm per seconds configurable
Delay: 6 seconds
Record Modes: Manual or automatic configurable
Battery
Type: Rechargeable lithium ion,10.8 Vdc, 8.61 Ah, 93 Wh
Capacity: At least 4 hours of continuous monitoring with ECG, SpO2, CO2, three invasive pressure channels, two temperature channels, NIBP measurements every 15 minutes, wireless communication disabled, display brightness set to 10%, and volume set to medium (with a new, fully charged battery operating at room temperature)
Weight: 1.5 lb.
Recharge Time: 5 hours to fully charge, max. When charged by Zenix
AC Power Adapter: 100 240 VAC 50-60 Hz, 160 VA
External Data Transfer
Data Transfer Port: USB
(Type A and C), Wi-Fi, Bluetooth
Monitor Port: Video Out
Wi-Fi: 802.11 a, b, g, n and ac
Bluetooth: 5.0
TRAINING:
Must provide training for the rapid response teams, nurse educators and bio-med.
DELIVERY:
Delivery shall be within 90 days after receipt of contract. Deliver equipment to:
Wilmington VA Medical Center (Attn: Shayonna Smith)
1601 Kirkwood Hwy
Wilmington, DE 19805-4917
POINT OF CONTACT (POC):
The identified POC for this acquisition is:
Shayonna Smith, Administrative Specialist, Patient Care Services
Email:shayonna.smith@va.gov
Phone Number: (302) 994-2511 ext. 34853 Work Cell: (302) 588-3289
Government personnel, other than the Contracting Officer, have the authority to change or alter these requirements. The POC shall clarify technical points or supply relevant technical information, but no requirements in this scope of work may be altered as a sole result of such verbal clarification.
REQUEST FOR INFORMATION INSTRUCTIONS:
The information identified above is intended to be descriptive, not restrictive and to indicate the quality of the supplies/services that will be satisfactory. It is the responsibility of the interested source to demonstrate to the government that the interested parties can provide the supplies/services that fulfill the required specifications mentioned above.
Responses to this RFI should include company name, address, point of contact, phone number, and point of contact e-mail, UEI Number, Cage Code, size of business pursuant to North American Industrial Classification System (NAICS) 334510 Please answer the following questions:
Please indicate the size status and representations of your business, such as but not limited to: Service-Disabled Veteran Owned Small Business (SDVOSB), Veteran Owned Small Business (VOSB), Hubzone, Woman Owned Small Business (WOSB), Large Business, etc.)?
Is your company considered small under the NAICS code identified under this RFI?
Are you the manufacturer, distributor, or an equivalent solution to the items being referenced above?
If not, can you provide additional information shown below. This is to confirm compliance with the non-manufacturer rule IAW 13 CFR 121.406(b) Nonmanufacturers.
Does your company exceed 500 employees;
Primarily engaged in the retail or wholesale trade and normally sells the type of item being supplied;
Take ownership or possession of the item(s) with its personnel, equipment or facilities in a manner consistent with industry practice;
If you are a large business, do you have any designated distributors? If so, please provide their company name, telephone, point of Contact and size status (if available).
If you re a small business and you are an authorized distributor/reseller for the items identified above or an equivalent solution, do you alter; assemble; modify; the items requested in any way? If you do, state how and what is altered; assembled; modified?
Must have an authorized distributor letter from the original equipment manufacture.
If you intend to subcontract any work on this contract, what portion of the total cost will be self-performed/will be performed by your organization? Please provide estimated detailed percentage breakdowns related to subcontracted work and completion of job.
Does your company have an FSS contract with GSA or the NAC or are you a contract holder with any other federal contract? If so, please provide the contract number.
If you are an FSS GSA/NAC contract holder or other federal contract holder, are the items/solution you are providing information for available on your schedule/contract?
General pricing of your products/solution is encouraged. Pricing will be used for the purpose of market research only. It will not be used to evaluate for any type of award.
Please submit your capabilities in regard to the salient characteristics detailed above and any information pertaining to equal to items to establish capabilities for planning purposes?
Please review salient characteristics/statement of work (if applicable) and provide feedback or suggestions. If none, please reply as N/A.
Equipment manufacture, Country of origin
Please provide your UEI number.
This RFI will be conducted in accordance with Federal Acquisition Regulation (FAR) Part 12. Telephone responses will not be accepted. Responses must be received via e-mail to andrew.taylor3@va.gov no later than, 4:00 PM Eastern Time (ET) on May 6, 2026. This notice will help the VA in determining available potential sources only. Do not contact VA Medical Center staff regarding this requirement, as they are not authorized to discuss this matter related to this procurement action.
All questions will be addressed by the Contracting Officer, Andrew Taylor. Questions or responses will include the Source Sought number 36C24526Q0546 in the subject line. All firms responding to this Request for Information are advised that their response is not a request for proposal, therefore will not be considered for a contract award.
Any interested business concern must submit a no longer that 10-page capability statement addressing its ability to meet the requirements listed above to Capability Statement shall include a statement regarding how the business concern is engaged in wholesale trade and normally sells the item being supplied and how the business concern will take ownership of the items being delivered. Vendors shall also provide certification, evidence that they are an authorized distributor of the Original Equipment Manufacturer.
If a solicitation is issued, information will be posted on the SAM web site for all qualified interested parties at a later date and interested parties must respond to the solicitation to be considered for award. This notice does not commit the government to contract for any supplies or services. The government will not pay for any information or administrative cost incurred in response to this Request for Information. Information will only be accepted in writing by e-mail to Contracting Officer at andrew.taylor3@va.gov include 36C24426Q0546 in the subject line.
DISCLAIMER
This RFI is issued solely for information and planning purposes only and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. Responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.
End of Document
Â